Product Recall for product correction - James Leckey Design Mygo Stander

12 July 2016

Product Recall for product correction - James Leckey Design Mygo Stander

The Australian Competition & Consumer Commission/Product Safety Australia has issued a recall notice for James Leckey Design Mygo Standers, supplied by Otto Bock Australia, with the following identifying features:

Mygo Stander Size 1

Product code:143-1600 Serial number range: LEC/087023 to LEC/108416 Month of manufacture: January 2012 to August 2013  

Mygo Stander Size 2

Product code:143-2600
Serial number range:LEC/087023 to LEC/108416
Month of manufacture: January 2012 to August 2013

The recall notice notes the design of the equipment' angle adjustment mechanism "...allows the user to lock out the frame's safety pull pin and adjust the angle using one hand. If the operator's hand slips off the turning handle when a patient is in the frame, the handle could 'self-wind' and spin rapidly as the stander returns to the upright position" and that "there is no risk to the patient in the frame, there is a small risk that the fast spinning handle could injure the operator if they try to take hold of it"

We have been advised that dealers will be contacting clients directly, to arrange for the product correction. Clients can contact Otto Bock Australia on 03 9439 9917.

Further details are available from Product Safety Australia.